Work

Ion motion in the rectangular wave quadrupole field and digital operation mode of a quadrupole ion trap mass spectrometer

Year: 2005
Type: article
Abstract: A quadrupolar electric field driven by a rectangular wave voltage can be used for mass-selective storage and analysis. The ion motion in such an electric field is derived, and the stability of ions is presented in the a-q diagram that is commonly used for sinusoidal wave quadrupole mass spectrometry in association with the solution of the Mathieu equation. The pseudo-potential well is discussed in an approximation that leads to the relation of secular frequency to operating parameters. A scheme for a digital ion trap mass spectrometer is described, based on this theory. An ion optics simulation was performed to check the theory of resonant ejection, and to prove the feasibility of the mass scan method for a practical ion trap of such geometry.
Language: English
Subfield: Spectroscopy
Field: Chemistry
Open Access status: bronze
OpenAlex ID: W1963995338
FWCI: 1.026
Cites: 4
Cited by: 59
Related to: 9