Signal quality improvement using cylindrical confinement for laser induced breakdown spectroscopy
Year: 2013
Type: article
Abstract: In our previous work, we found that there was great potential to improve the pulse-to-pulse signal repeatability using a moderate cylindrical cavity confinement. However, the improvement was achieved only with certain experimental parameters; while under other conditions, there was no improvement or even worse repeatability. In the present work, the experimental configuration was redesigned and unexpected uncertainty from the variation of the laser and cavity alignment and the laser ablated aerosols were avoided. With these two improvements, we demonstrated that the cavity can always increase the signal repeatability. In addition, image taken by ICCD verified that the confinement improved the stability of the plasma morphology as expected.
Source: Optics Express
DOI: 10.1364/oe.21.015974
Authors: Zongyu Hou, Zhe Wang, Jianmin Liu, Weidou Ni, Zheng Li
Language: English
Subfield: Spectroscopy
Field: Chemistry
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $2,270
OpenAlex ID: W1985386710