Precision synchronous polarimeter with linear response for the measurement of small rotation angles
Year: 2000
Type: article
Abstract: A synchronous polarimeter was set up for the measurement of small rotation angles of the polarization plane of light. The polarimeter is based on a polarizer-Faraday modulator-analyzer structure with a synchronous detection scheme, which produces a linear system response. The theoretical background is studied, and the system performance is investigated experimentally. We achieved an accuracy of the order of 10(-4) deg, or 5 mg/dl of glucose in a 1-cm light path.
Source: Applied Optics
DOI: 10.1364/ao.39.002601
Institution: Universidad de la República de Uruguay
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W1990395960