Development and Evaluation of a Novel In-Clinic Automated Hematology Analyzer, ProCyte Dx, for Canine Erythrocyte Indices, Leukogram, Platelet Counts and Reticulocyte Counts
Year: 2013
Type: article
Abstract: A novel hematology analyzer for small animal medicine, ProCyte Dx, was developed from combination of the fluorescence laser flow cytometry and laminar flow impedance technologies, and its accuracy was evaluated by comparing with the conventional impedance-based hematology analyzer, pocH-100iV Diff, or microscopic manual cell counting methods with staining blood smears in the canine blood. Blood samples of 59 dogs were hematologically analyzed and compared by Pearson's correlation coefficients. Analyses between the two analyzers showed excellent correlation in RBC (r=0.998), HGB (r=0.999), HCT (r=0.998), MCV (r=0.994), MCH (r=0.974), MCHC (r=0.906), WBC (r=0.998) and PLT (r=0.993). Analyses between ProCyte Dx and microscopic manual counting results showed excellent correlation in neutrophils (r=0.920), lymphocytes (r=0.913) and reticulocyte percentages (r=0.924), good correlation in eosinophils (r=0.815) and reticulocyte numbers (r=0.850) and fair correlation in monocytes (r=0.770). The present study indicates that ProCyte Dx is acceptably accurate and can be a powerful tool for canine clinical medicine.
DOI: 10.1292/jvms.13-0264
Authors: Yasuhito FUJINO, Yoichi NAKAMURA, Hideaki MATSUMOTO, Kenjiro FUKUSHIMA, Masashi TAKAHASHI, Koichi OHNO, Hajime TSUJIMOTO
Institutions: The University of Tokyo, Sysmex (Japan)
Language: English
Subfield: Medical Laboratory Technology
Field: Health Professions
Domain: Health Sciences
Open Access status: hybrid
OpenAlex ID: W2000437242