Work

Low-cost conversion of a coaxial nozzle arrangement into a stationary low-temperature attachment

Year: 1996
Type: article
Abstract: A stationary nozzle accessory designed for the Siemens/Nicolet LTII low-temperature device with standard ϕ-coaxial nozzle arrangement is described. The attachment is easy to assemble and its compatibility allows interchanging between both systems in less than 1 h, if required. Implementation of the stationary nozzle reduces the consumption of liquid nitrogen by more than 40% and extends the temperature range available for data collection by 60 K in the lower temperature range. Details of construction are presented, and a comparison with the ϕ-coaxial nozzle is discussed.
Language: English
Open Access status: bronze
OpenAlex ID: W2002241371
FWCI: 8.228
Cites: 9
Cited by: 196
Related to: 10