Year: 2002
Type: article
Abstract: We present the multiplexed readout of two gamma-ray microcalorimeters made from transition-edge sensors. We use a frequency-domain multiplexing scheme in which each sensor is biased at an identifying frequency. We show that the energy resolution of the sensors is unaffected by multiplexing, and that crosstalk between the sensors is negligible. Our results indicate the feasibility of multiplexing 30 sensors or more to one readout line.
Source: Applied Physics Letters
DOI: 10.1063/1.1489486
Language: English
Subfield: Electrical and Electronic Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green