Calculated effective atomic number and Kerma values for tissue-equivalent and dosimetry materials
Year: 1971
Type: article
Abstract: The calculated value of the effective atomic number (z tau ) of 7.23 for lithium tetraborate (Li 2 B 4 O 7 ) agrees closely with the value of 7.22 for soft tissue of composition C 5 H 40 O 18 N. For excised muscle specimens the experimental value (7.42) obtained by Spiers for z tau is very nearly equal to the corresponding calculated value (7.46) for the composition (ICRU) of striated muscle. A similar comparison shows that the composition for bone given by Woodard with a higher content of calcium and phosphorus gave a value of z tau =13.3, which is in closer agreement with the measured value of 13.8 obtained by Spiers. The calculated values of z tau and kerma per roentgen for lithium tetraborate and lithium fluoride for photon energies below 1 MeV show that lithium borate (Li 2 B 4 O 7 ) is a better substitute for soft tissue than lithium fluoride.
Source: Physics in Medicine and Biology
Author: C A Jayachandran
Institution: Christian Medical College, Vellore
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: bronze
OpenAlex ID: W2010366245