Recommended design parameters for thermal conductivity probes for nonfrozen food materials
Year: 1996
Type: article
Abstract: This paper analyzes the various design parameters of the thermal conductivity (k) probe and makes design recommendations for applications for nonfrozen food materials. The k probe is a simplified application of the line-heat source theory, but this simplification contributes to the instrument error which can be minimized by paying particular attention to the size of the k probes and materials of construction. As the diameter of the k probe decreases, its accuracy increases; however, small k probes are difficult to make. Therefore, it is recommended that users design their k probes using the highest acceptable error for their intended applications.
Source: Journal of Food Engineering
Authors: E.G. Murakami, V.E. Sweat, S.K. Sastry, E. Kolbe, K. Hayakawa, A. Datta
Language: English
Subfield: Computational Mechanics
Field: Engineering
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $4,350
OpenAlex ID: W2011302536