Dosimetry of 192Ir wires for LDR interstitial brachytherapy following the AAPM TG-43 dosimetric formalism
Year: 2001
Type: article
Abstract: Implementation of the AAPM Task Group 43 dosimetric formalism for 192Ir wires used as interstitial sources in low dose-rate (LDR) brachytherapy applications is investigated. Geometry factors, dose-rate constant values, radial dose functions, and anisotropy functions, utilized in this formalism, were calculated for various lengths of all commercially available wire source designs by means of a well-established Monte Carlo simulation code and an improved modification of the Sievert integral method. Results are presented in the form of look up tables that allow interpolation for dose-rate calculations around all practically used wire lengths, with accuracy acceptable for clinical applications.
Source: Medical Physics
DOI: 10.1118/1.1339885
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green