Work

Dosimetry of 192Ir wires for LDR interstitial brachytherapy following the AAPM TG-43 dosimetric formalism

Year: 2001
Type: article
Abstract: Implementation of the AAPM Task Group 43 dosimetric formalism for 192Ir wires used as interstitial sources in low dose-rate (LDR) brachytherapy applications is investigated. Geometry factors, dose-rate constant values, radial dose functions, and anisotropy functions, utilized in this formalism, were calculated for various lengths of all commercially available wire source designs by means of a well-established Monte Carlo simulation code and an improved modification of the Sievert integral method. Results are presented in the form of look up tables that allow interpolation for dose-rate calculations around all practically used wire lengths, with accuracy acceptable for clinical applications.
Language: English
Subfield: Radiation
Open Access status: green
OpenAlex ID: W2012755733
FWCI: 4.584
Cites: 19
Cited by: 23
Related to: 10