Calibration and performance of the E706 lead and liquid-argon electromagnetic calorimeter
Year: 1998
Type: article
Abstract: We report on the calibration and performance of a large lead liquid-argon electromagnetic calorimeter used by experiment E706 at Fermilab. The reconstructed π 0 mass was used to calibrate the energy response of the calorimeter. The systematic uncertainty in the linearity and uniformity of the mean-energy response of the calorimeter after calibration was found to be less than ±0.5% for the sample of data analyzed. Detector characteristics, including sampling fluctuations and position resolution, are discussed.
Authors: L. Apanasevich, J. Bacigalupi, W. Baker, G. Ballocchi, M. Begel, R. Benson, S. Blusk, C. Bromberg, D. Brown, P. Chang, B. Choudhary, W.H. Chung, L.de Barbaro, W. DeSoi, W. Długosz, J. Dunlea, E. Engels, G. Fanourakis, T. Ferbel, J. Ftacnik, D. Garelick, G. Ginther, M. Glaubman, P. Gutierrez, K. Hartman, J. Huston, C. Johnstone, V. Kapoor, J. Kuehler, A. Lanaro, C. Lirakis, F. Lobkowicz, P. Lukens, S. Mani, J. Mansour, A. Maul, R. Miller, B.Y. Oh, G. Osborne, D. Pellett, E. Prebys, R. Roser, A. Sinanidis, P. Shepard, R. Shivpuri, D. Skow, P. Slattery, L. Sorrell, D. Striley, W. Toothacker, N. Varelas, D. Weerasundara, J.J. Whitmore, G. Wu, T. Yasuda, C. Yosef, M. Zieliński, V. Zutshi