Work

X-ray focusing with lobster-eye optics: a comparison of theory with experiment

Year: 1996
Type: article
Abstract: We report an experimental investigation and comparison with simulation of the x-ray focusing of a flat, square profile microchannel plate. We use x rays with an energy of ~1.5 keV from a laser-produced plasma. The images were recorded with x-ray film. We find the focal structure to be consistent with theoretical expectations. The angular resolution of the focus is 0.96 mrad, which is a major improvement over previous results. The measured peak intensity gain is 27 ± 4, which is ~33% of that for a perfect optic.
Language: English
Subfield: Instrumentation
Open Access status: green
OpenAlex ID: W2015406846
FWCI: 5.362
Cites: 19
Cited by: 47