Year: 1965
Type: report
Abstract: A neutron detector which has application as both an area monitoring instrument and a criticality dosimeter is under development at Argonne National Laboratory. The detector simultaneously measures both dose equivalent and fluence from an exposure to fast neutrons. The moderator consists of a 12-in. -diameter, aluminum-encased paraffin sphere. The neutron sensors are nine thermal activation foils located within the moderator. The neutron fluence is determined by activation of six foils symmetrically located one inch below the surface of the moderator. For this foil array, the summed activity is nondirectional and proportional to the fluence within 10 percent over the investigated energy range from 20 keV to 2.3 MeV. The neutron dose equivalent is determined from activation of three symmetrically interlocked foils at the center of the moderator, and the summed activity is proportional to the dose equivalent within 75 percent over the same energy range. A response correction technique is described which, for monoenergetic neutrons, brings the dose equivalent to within about 10 percent. The average sensitivity of the monitor over the stated energy range, using indium foils, is 76 cpm per mrem/hr and 25 cpm per n/cm2-sec at the end of one half-life exposure at uniform radiation level.
DOI: 10.2172/4616320
Authors: Robert Dvorak, Norman Dyer
Institution: Argonne National Laboratory
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green