Work

Quality assurance: Recommended guidelines for safe heating by capacitive-type heating technique to treat patients with metallic implants

Year: 2013
Type: article
Abstract: This article is a redissemination of the previous Japanese Quality Assurance Guide guidelines. Specific absorption rate and temperature distribution were investigated with respect to various aspects including metallic implant size and shape, insertion site, insertion direction, blood flow and heating power, and simulated results were compared with adverse reactions of patients treated by radio frequency capacitive-type heating. Recommended guidelines for safe heating methods for patients with metallic implants are presented based on our findings.
Language: English
Open Access status: bronze
OpenAlex ID: W2037201708
FWCI: 0.622
Cites: 10
Cited by: 10