Abstract: BACKGROUND: Surgibone Unilab is prepared from bovine bone and contains hydroxyapatite and protein. It is supposed to be immunogenically inert but the protein could be antigenic in man. PATIENTS AND METHODS: We followed 27 patients for an average of 2.5 (1-5) years, all of whom had received Surgibone mixed with autograft to fill in defects in the acetabulum and the proximal femur in revision hip surgery. RESULTS: In 17 patients, there was apparently complete incorporation of the bone graft within 6 months. In 3 of these patients, the graft was incorporated after 3 months. In 3 patients, however, there was no incorporation of the graft as late as 3 years after the operation. 3 other patients appeared to have a type of graft rejection (pseudoinfection). 1 other patient suffered MRSA deep infection of the prosthesis which resulted in removal of the implants 1 month postoperatively. INTERPRETATION: Use of Surgibone xenograft in revision hip surgery, even in combination with autograft, resulted in failure and the need for rerevision in at least one quarter of the cases studied.
Institution: Epsom Hospital
Language: English
Subfield: Surgery
Field: Medicine
Open Access status: gold
Estimated APC paid: $1,779
OpenAlex ID: W2071944766
FWCI: 1.619
Cites: 16
Cited by: 49
Related to: 9