Work

Optimization of THz wave coupling into samples in DNP/NMR spectroscopy

Year: 2010
Type: conference-paper
Abstract: High power millimeter wave and terahertz sources are used in DNP/NMR spectroscopy to greatly enhance the NMR signal. A key issue is the efficient coupling of the source power to the sample. We present HFSS calculations showing that the MIT 250 GHz DNP/NMR system provides a B1value of 13 μT/W1/2. The simulation shows the importance of the diffraction of radiation passing through the RF coil, leading to non-uniform illumination of the sample. A simple single pass model yields insight into the origin of the observed B1value. Methods of optimizing the B1value are described.
Language: English
Subfield: Spectroscopy
Field: Chemistry
Open Access status: green
OpenAlex ID: W2080831706
FWCI: 2.305
Cites: 8
Cited by: 2
Related to: 10