Perturbation correction factors for some ionization chambers commonly applied in electron beams

Year: 1991
Type: article
Abstract: Values for the fluence perturbation correction factor were determined in a number of electron beams for the commonly applied PTW/Markus plane-parallel chamber and the cylindrical NE 2571 Farmer chamber. These data were determined relative to the NACP plane-parallel chamber as a function of the mean energy at depth, E z , calculated according to the method recommended in modern dosimetry protocols. For the cylindrical chamber the results are in agreement with the data recommended in these protocols. For the PTW/Markus chamber a small but significant fluence perturbation correction is necessary: up to about 3% for a value of E z =2 MeV. In order to check the value of these fluence perturbation correction values, dose measurements were performed ionometrically and compared with values determined with the Fricke dosimetry system.
Language: English
Subfield: Radiation
Open Access status: bronze
OpenAlex ID: W2081441925
FWCI: 2.307
Cites: 18
Cited by: 33
Related to: 10