Weighted FBP—a simple approximate 3D FBP algorithm for multislice spiral CT with good dose usage for arbitrary pitch
Year: 2004
Type: article
Abstract: A new 3D reconstruction scheme, weighted filtered backprojection (WFBP) for multirow spiral CT based on an extension of the two-dimensional SMPR algorithm is described and results are presented. In contrast to other 3D algorithms available, the algorithm makes use of all available data for all pitch values. The algorithm is a FBP algorithm: linear convolution of the parallel data along the row direction followed by a 3D backprojection. Data usage for arbitrary pitch values is maintained through a weighting scheme which takes into account redundant data. If proper row weighting is applied, the image quality is superior to the image quality of the SMPR algorithm.
Source: Physics in Medicine and Biology
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W2088596249
