Work

Reducing Measurement Uncertainty of Instruments Based on the Phenomenon of Surface Plasmon Resonance

Year: 2013
Type: article
Abstract: Considered in this paper are mechanisms of temperature influence on the results of optical measurements based on the surface Plasmon resonance phenomenon. It has been ascertained that the error of measurement results is related with the temperature influence on the refraction index of substances under study as well as on elements of optical setup in measuring equipment. The authors have experimentally demonstrated the influence of thermal stabilization of measuring equipment on dispersion of results obtained in optical measurements.
Language: English
Open Access status: bronze
OpenAlex ID: W2133554640
FWCI: 2.206
Cites: 24
Cited by: 13