Reducing AC impedance measurement errors caused by the DC voltage dependence of broadband high-voltage bias-tees

Year: 2007
Type: conference-paper
Abstract: During the AC impedance characterization of devices, from the kHz-range up to the GHz-range, accuracy can be lost when a DC voltage is applied. Commercial high-voltage broadband bias-tees are often voltage-dependent, which can cause inaccuracies at low frequencies. A calibration technique with applied bias significantly improves the measurement accuracy. Additionally, a bias-tee has been developed with a voltage-independent capacitor, suitable for DC voltages up to 500 V showing excellent performance up to several gigahertz. PIN diode limiters protect the measurement equipment from damage in case of a device breakdown.
Language: English
Open Access status: green
OpenAlex ID: W2137736364
FWCI: 1.004
Cites: 12
Cited by: 11
Related to: 10