Ventricular Tachycardia Recognized as Ventricular Fibrillation by an Implantable Cardioverter Defibrillator, Due to Double Counting of the QRS Complex
Year: 2013
Type: article
Study design: Case Report
Abstract: A case of ventricular tachycardia misdiagnosed as ventricular fibrillation due to double counting of the tachycardia’s wide QRS complex in an implantable cardioverter defibrillator (ICD) recipient is presented. As a result, an inappropriate shock was delivered by the device instead of anti-tachycardia pacing. The reprogramming of the detection criteria in a less sensitive manner solved the problem.
DOI: 10.2015/hc.v8i1.469
Institutions: Hellenic Red Cross, Red Cross Hospital
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: diamond
OpenAlex ID: W2139306357