Work

Intercomparison of erythemal broadband radiometers calibrated by seven UV calibration facilities in Europe and the USA

Year: 2008
Type: article
Abstract: . A bi-lateral intercomparison of erythemal broadband radiometers was performed between seven UV calibration facilities. The owners calibrations were compared relative to the characterisation and calibration performed at PMOD/WRC in Davos, Switzerland. The calibration consisted in the determination of the spectral and angular response of the radiometer, followed by an absolute calibration performed outdoors relative to a spectroradiometer which provided the absolute reference. The characterization of the detectors in the respective laboratories are in good agreement: The determination of the angular responses have deviations below ±4% and the spectral responses agree within ±20%. A "blind" intercomparison of the erythemally weighted irradiances derived by the respective institutes and PMOD/WRC showed consistent measurements to within ±2% for the majority of institutes. One institute showed slightly larger deviation of 10%. The differences found between the different instrument calibrations are all within the combined uncertainty of the calibration.
Language: English
Open Access status: gold
Estimated APC paid: $1,725
OpenAlex ID: W2139647789
FWCI: 5.83
Cites: 11
Cited by: 27
Related to: 10