Performance of a nanometer resolution BPM system

Year: 2006
Type: report
Abstract: International Linear Collider (ILC) interaction region beam sizes and component position stability requirements will be as small as a few nanometers. It is important to the ILC design effort to demonstrate that these tolerances can be achieved – ideally using beam-based stability measure- ments. It has been estimated that RF cavity beam position monitors (BPMs) could provide position measurement res- olutions of less than one nanometer and could form the ba- sis of the desired beam-based stability measurement. We have developed a high resolution RF cavity BPM system. A triplet of these BPMs has been installed in the extrac- tion line of the KEK Accelerator Test Facility (ATF) for testing with its ultra-low emittance beam. The three BPMs are rigidly mounted inside an alignment frame on variable- length struts which allow movement in position and angle. We have developed novel methods for extracting the posi- tion and tilt information from the BPM signals including a calibration algorithm which is immune to beam jitter. To date, we have been able to demonstrate a resolution of ap- proximately 20 nm over a dynamic range of +/- 20 microns. We report on the progress of these ongoing tests.
Language: English
Open Access status: green
OpenAlex ID: W2149677669
Cites: 1