Pacemaker syndrome and pseudo-ventricular high threshold after dual-chamber pacemaker replacement
Year: 2010
Type: article
Study design: Case Report
Abstract: We report a 60-year-old male patient who presented with pacemaker syndrome. The patient had recent dual-chamber pacemaker replacement. Surface ECG and endocavitary electrograms were compatible with leads connection inversion. Device interrogation showed pseudo-ventricular high threshold. Dual-chamber pacemaker leads inversion in a pacer-dependent patient can lead to life-threatening complications. Industry technicians should not perform device technical assistance when unsupervised by physicians with expertise.
Source: EP Europace
Authors: A. Kossaify, N. Moussallem
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W2151206363