Work

A Validated RP HPLC Method for the Determination of Moprolol in Bulk and Tablet Dosage Form

Year: 2009
Type: article
Abstract: A validated reverse phase high performance liquid chromatographic method (RP-HPLC) has been developed for the estimation of moprolol in the pharmaceutical preparation using RPC 18 column. The mobile phase (Acetonitrile:Methanol:Water) was pumped at a flow rate of 1ml/min in the ratio of 15:50:35 and the eluents were monitored at 274nm at 30°C. The intra and interday precision was found to be less than 2% showing high precision of the assay method. The % recovery of the method was more than 99% and RSD did not exceed 2% indicating high degree of accuracy of the proposed HPLC method. The %RSD for the robustness testing was also less than 2%.The proposed HPLC method can be used for the estimation of moprolol in tablet dosage forms.
Language: English
Field: Chemistry
Open Access status: diamond
OpenAlex ID: W2201825541
FWCI:
Cites:
Cited by: 5
Related to: 10