Year: 1934
Type: article
Abstract: In order to test the ice point as a suitable temperature at which to measure pure-metal resistance standards, an investigation has been made of its repro- ducibility.The results obtained show that the ice point may be readily repro- duced to a few ten thousandths of a degree centigrade.This result was arrived at by intercomparing different ice baths, and by comparing the ice point with the triple point of water.
DOI: 10.6028/jres.012.031
Author: J.L. Thomas
Institution: United States Department of Commerce
Language: English
Subfield: Instrumentation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: hybrid
OpenAlex ID: W2331007288