Investigation of TLD-700 energy response to low energy x-ray encountered in diagnostic radiology

Year: 2016
Type: article
Abstract: The aim of thiswork is to study the energy dependence of thermoluminescent dosimeter (TLD-700) for low energy X-ray beams encountered in conventional diagnostic radiology. In the first step, we studied some characteristics (reproducibility and linearity) of TLD-700 chips using a 137Cs source, and selected TLD chips with reproducibility better than 2.5%. Then we determined TLD-700 energy response for diagnostic radiology X-ray qualities, and investigated its influence on air kerma estimate. A maximum deviation of 60% can be obtained if TLDs are calibrated for 137Cs radiation source and used in diagnostic radiology fields. However, this deviation became less than 20% if TLDs chips are calibrated for the reference x-ray radiation quality RQR5 (recommended by the IEC 61267 standard). Consequently, we recommend calibrating this kind of TLDdetector with RQR5 diagnostic radiology X-ray quality. This method permits to obtain a good accuracy when assessing the entrance dose in diagnostic radiology procedures.