Surface Changes Of WaveOne™ and Reciproc® Instruments after Using Three Times for Preparation of Simulated Curved Canals with and without Glide Path.

Year: 2019
Type: article
Abstract: Introduction: when used in the preparation of simulated curved canals with and without glide path (generated with the Pathfile™ system), after the first, second, and third uses. Materials and methods: instruments, respectively. Each instrument was used three times; after each use, each instrument was analyzed by using scanning electron microscopy (cervical, middle, and apical thirds of the instrument) in order to characterize the occurrence of changes (fracture, twist, and crack). Data were described using means and standard deviations. We used generalized linear models to compare differences between factors (region, manufacturer, glide path, and number of uses). SPSS-15 software was used, with a significance level of 5%. Results: instruments did not exhibit superficial changes, such as fractures and/or twists. Conclusion: instruments exhibited superior performance, compared with WaveOne™ instruments, particularly when glide path with the PathFile™ system was used; both instruments may be used, safely, three times to prepare curved canals.