Use of sensor‐integrated pump therapy to reduce hypoglycaemia in people with Type 1 diabetes: a real‐world study in the UK

Year: 2019
Type: article
Study design: Observational Study
Abstract: AIMS: To assess the efficacy of insulin pumps with automated insulin suspension systems in a real-world setting. METHODS: by people (n=920) with Type 1 diabetes using the MiniMed Paradigm Veo system and the MiniMed 640G system (Medtronic International Trading Sàrl, Tolochanez, Switzerland) with SmartGuard technology, with or without automated insulin suspension enabled, between February 2016 and June 2018. Users with ≥15 days of sensor data and ≥70% sensor-wear time were classified as sensor-augmented pump alone, sensor-integrated pump with low glucose suspend enabled or sensor-integrated pump with predictive low glucose management enabled. RESULTS: percentile) system use was 161 (58-348) days. The median time spent with sensor glucose values ≤3 mmol/l was 0.8 (0.3-1.7)% in the sensor-augmented pump group, 0.3 (0.1-0.7)% in the sensor-integrated pump with low glucose suspend group, and 0.3 (0.1-0.5)% in the sensor-integrated pump with predictive low glucose management group. In individuals switching from sensor-augmented pump to sensor-integrated pump with low glucose suspend (n=31), there were significant reductions in the monthly rate of hypoglycaemic events <3 mmol/l (rate ratio 0.63, 95% CI 0.45-0.89; P=0.009) and in the percentage of time with glucose values ≤3 mmol/l [sensor-augmented pump: 0.63% (95% CI 0.34-1.29), sensor-integrated pump with low glucose suspend: 0.33% (95% CI 0.16-0.64); P=0.001]. The monthly rate of hypoglycaemic events decreased further in individuals (n=139) switching from sensor-integrated pump with low glucose suspend to sensor-integrated pump with predictive low glucose management [rate ratio 0.82 (95% CI 0.69-0.98); P<0.0274]. Similar results were seen for events <3.9 mmol/l. There was no difference in median time spent in target glucose range. CONCLUSION: Real-world UK data show that increasing automation of insulin suspension reduces hypoglycaemia exposure in people with Type 1 diabetes.
Language: English
Field: Medicine
Open Access status: bronze
Funder: Medtronic
OpenAlex ID: W2946917435
FWCI: 1.713
Cites: 27
Cited by: 22
Related to: 10