Year: 2019
Type: article
Abstract: The aim of the study was to investigate the clinical feasibility of whole-brain radiation therapy without a thermoplastic mask. Positioning and intra-fractional motion monitoring were performed using optical surface scanning. The motion threshold was 3 mm/3 degrees. The group mean vector deviation was 1.1 mm. The roll was larger compared to pitch and rotation. Two patients out of 30 were not able to lie still. All other patients completed their treatment successfully without a mask. With a probability of success of 93%, we concluded that irradiation without a mask is a clinically feasible method.
Institutions: Instituut Verbeeten, Tilburg University
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $1,538
Funder: Verbeeten Fonds
OpenAlex ID: W2964337129