Using Electrode Made of Carbon Nanotubes and Bismuth Oxide for the Determination of Metal Concentration by Anodic Stripping Voltammetry
Year: 2019
Type: article
Abstract: We have successfully manufactured a new electrode modified with bismuth oxide (Bi2O3) using carbon nanotubes (CNTs). The electrode was fabricated to detect cadmium (Cd), lead (Pb), and indium (In) by differential pulse anodic stripping voltammetry (DP-ASV). The electrode surface was studied by scanning electron microscopy (SEM), and the reduction and oxidation processes were studied by cyclic voltammetry (CV) techniques. Operational parameters such as electrode size, bismuth concentration, and electrolytic background were optimized. The DP-ASV method used fabricated electrodes with a linear response range from 1.5–20 μg·L−1with Cd(II) and Pb(II) and 2.5–20 μg·L−1with In(III); low detection limit (LOD) of 0.22 μg·L−1with Cd(II), 0.65 μg·L−1with In(III), and 0.26 μg·L−1with Pb(II); and good repeatability with relative standard deviations (RSD) of 2.65%, 2.51%, and 3.34% with Cd(II), Pb(II), and In(III), respectively (n = 8). The electrode can be used to test the content of Cd(II), In(III), and Pb(II) in water.
Source: Journal of Chemistry
DOI: 10.1155/2019/6170967
Language: English
Subfield: Electrochemistry
Field: Chemistry
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $2,160
OpenAlex ID: W2972001127