Bony Resection Outliers: Insights from 10,144 Clinical Cases using A Contemporary Computer-Assisted Total Knee Arthroplasty System
Year: 2019
Type: conference-paper
Study design: Observational Study
Abstract: This study accessed the alignment outliers of intraoperatively measured bony resection during total knee arthroplasty on 10,144 cases performed using a modern CAOS system. The impacts from geographic regions, surgeon’s adoption of the technology (learning or proficient phases), and historical progression of the CAOS application (software versions) were evaluated. The comprehensive analysis demonstrated that the CAOS system is a robust and accurate solution to assist the surgeons to achieve his/her surgical resection goals across its application history.
Source: EPiC series in health sciences
DOI: 10.29007/bwjk
Institution: Exactech (United States)
Language: English
Subfield: Surgery
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W3020143764