Development of Ultrahigh Resolution Objective Lens Enabling High Analytical Sensitivity

Year: 2020
Type: article
Abstract: Spatial resolution of transmission electron microscope (TEM) with spherical aberration corrector has been improved to be 40 pm [1], and it enables ordinary observation of materials samples at atomic resolution.Recently, such atomic resolution has become required not only for high resolution imaging at high accelerbut observations with a variety of methods and conditions.Among these various observations, the combination of scanning transmission electron microscopy (STEM) and energy dispersive X-ray spectroscopy (EDS) is typically popular and requested from huge number of researchers, because it gives fundamental properties of materials.The EDS sensitivity improves, when the EDS detector set at closed position to the specimen and/or the detection area of the EDS detector is large.However, the space near the specimen is limited by the objective pole-piece.Therefore, we tried to optimize the position of the detector by redesigning the objective lens polepiece of ultrahigh resolution to achieve the high resolution and EDS sensitivity simultaneously.The shape of the objective pole-piece for high resolution TEMs is generally thick to avoid the magnetic saturation.And the gap between pole-pieces is narrow to make sharp magnetic field distribution, which results in the reduction of the chromatic aberration coefficient.