Work

Validation of combinatorial evaluation of strain-gauge amplifier linearity

Year: 2020
Type: article
Abstract: This paper describes a validation of a combinatorial calibration technique based calibration procedure for strain-gauge amplifier calibration. The deviation between two strain-gauge amplifiers at calibration using the combinatorial technique is compared to the deviation between the same two amplifiers at calibration on a force calibration machine as a transducer-amplifier chain under unchanged conditions, serving as an amplifier linearity comparator. This enables the validation to be confirmed with a suitable expanded measurement uncertainty of 3 nV/V.
Source: ACTA IMEKO
Author: M. Hiti
Language: English
Subfield: Instrumentation
Open Access status: diamond
OpenAlex ID: W3119852575
FWCI:
Cites: 5
Cited by:
Related to: 10