Work

Development and Validation of a Ultra Flow Liquid Chromatography Method for the Assay of Boceprevir Using a Quality-by-Design Approach

Year: 2020
Type: article
Abstract: Quality by design guided. The assay method of Boceprevir is developed in accordance with ICH Q8(R2) guideline with due validation. .In this process, the Target analytical profile (TAP) of the drug was set and critical method parameters (CMP) were investigated by systematic risk assessment experimentation to control critical Quality Attributes (CQA). In this, A Cause Effect Risk Assessment Matrix with Control-Noise-Experiment (CNX) is used for identifying the high-risk variables i.e Percentage of Organic Modifier (% methanol), pH of the Buffer and flow rate of the mobile phase. The surface response methodology was applied to optimize the critical method parameters (CMP) as well as Critical Quality Attributes (CQA) to find out the Design space of the method. The Optimum assay method condition was mobile phase Acetate Buffer (50mM) pH 5.4: Methanol (11:89), Flow rate: 0.9 ml/min, Lambda Max: 207. The separation was achieved in the Eclip Plus C-18 column (250 × 4.6 mm, 5μm) at ambient temperature. The retention time of Boceprevir was found to be 4.2 min. The method evaluation was performed according to the (Q2R1) ICH guideline.
Language: English
Field: Chemistry
Open Access status: diamond
OpenAlex ID: W3144846397
FWCI:
Cites:
Cited by:
Related to: 10