Cavopulmonary Support Operating Criteria for Serving the Failing Fontan Population
Year: 2019
Type: preprint
Abstract: Fontan operation as the current ultimate palliation of single ventricledefects results in significant late complications eventually leading to a
failing circulation. It has been suggested that introducing a cavopulmonary
assist device to serve the function of the missing subpulmonary ventricle could
potentially stabilize the failing Fontan circulation. The objective of this
study is to identify the desired operating region for a cavopulmonary blood
pump that can offer a promising alternative treatment option for a wide range
of failing Fontan patients. By integrating numerical analysis and available
clinical information, the interaction of the cavopulmonary support (producing
head pressure rise from 2-20mmHg) via the IVC and full assist configurations
with a wide range of simulated adult failing patient cases was investigated;
with IVC and full assist corresponding to the inferior venous return or the
entire venous return, respectively, being routed through the device. We
identified the desired hydraulic operating regions for a cavopulmonary assist
device by clustering all head pressures and corresponding pump flows that
result in hemodynamic improvement for each simulated failing Fontan physiology.
The results presented here can serve as the performance criteria for designing
cavopulmonary assist devices as well as evaluating off-label use of
commercially available left-side blood pumps for failing Fontan cavopulmonary
support.