Work

Ppb-level gas detection using on-beam quartz-enhanced photoacoustic spectroscopy based on a 28 kHz tuning fork

Year: 2021
Type: article
Abstract: In this paper, an on-beam quartz-enhanced photoacoustic spectroscopy (QEPAS) sensor based on a custom quartz tuning fork (QTF) acting as a photoacoustic transducer, was realized and tested. The QTF is characterized by a resonance frequency of 28 kHz, ~15% lower than that of a commercially available 32.7 kHz standard QTF. One-dimensional acoustic micro resonator (AmR) was designed and optimized by using stainless-steel capillaries. The 28 kHz QTF and AmRs are assembled in on-beam QEPAS configuration. The AmR geometrical parameters have been optimized in terms of length and internal diameter. The laser beam focus position and the AmR coupling distance were also adjusted to maximize the coupling efficiency. For comparison, QEPAS on-beam configurations based on a standard QTF and on the 28 kHz QTF were compared in terms of H2O and CO2 detection sensitivity. In order to better characterize the performance of the system, H2O, C2H2 and CO2 were detected for a long time and the long-term stability was analyzed by an Allan variance analysis. With the integration time of 1 s, the detection limits for H2O, C2H2 and CO2 are 1.2 ppm, 28.8 ppb and 2.4 ppm, respectively. The detection limits for H2O, C2H2 and CO2 can be further improved to 325 ppb, 10.3 ppb and 318 ppb by increasing the integration time to 521 s, 183 s and 116 s
Language: English
Subfield: Spectroscopy
Field: Chemistry
Open Access status: gold
Estimated APC paid: $2,070
OpenAlex ID: W4200423007
FWCI: 6.481
Cites: 47
Cited by: 95
Related to: 10