Scanning Distance Influence on the Intraoral Scanning Accuracy—An In Vitro Study

Year: 2022
Type: article
Abstract: Intraoral scanners (IOS) have reached a point where their impact in the dental office cannot be denied. The distance between the tip of the IOS and the preparation may have implications on the accuracy of the digital model. The objective of this study was to evaluate the differences in accuracy between digital impressions in the scenario of different scanning distances. Twenty consecutive scans were performed at five predetermined distances: 5 mm, 10 mm, 15 mm, 20 mm and 23 mm by a single operator. The scanning distance of 10 mm displayed the best accuracy with an overall trueness value of 23.05 μm and precision value of 4.2 μm. The drawn conclusion was that increased scanning distances can decrease the accuracy of a digital impression.