Abstract: Among the major challenges of medicine today there are screening and early detection of tumors (since the adenoma stage) in order to prevent their degeneration into malignant cancer and/or metastases. In particular, the colorectal cancer shows a high curability rate, up to 90%, if identified when in its benign stage. The Protocol discussed here is proposed to implement the clinical validation of a device consisting of an array of chemoresistive gas sensors made of semiconductor materials, able of identifying the difference between fecal exhalation of healthy subjects and of subjects suffering from high-risk colorectal polyps or tumors. The tests are compared to the results of fecal occult blood test and colonoscopy as a gold standard.