Work

Pixelated silicon detector for radiation beam profile measurements

Year: 2023
Type: article
Abstract: A pixelated silicon detector, developed originally for particle physics experiments, was used for a beam profile measurement of a cobalt-60 (Co-60) irradiator in a water phantom. The beam profile was compared to a profile measured with a pinpoint ionization chamber. The differences in the pixel detector and pinpoint chamber relative profiles were within approximately 2% of profile maximum, and after calculating correction factors with Monte Carlo simulations for the pixel detector, the maximum difference was decreased to approximately 1% of profile maximum. The detector’s capability to measure pulse-height was used to record an electron pulse-height spectrum in water in the Co-60 beam, and the results agreed well with simulations.
Language: English
Subfield: Radiation
Open Access status: hybrid
Estimated APC paid: $2,910
OpenAlex ID: W4367692987
FWCI: 0.679
Cites: 20
Cited by: 4
Related to: 10