Work

Interlaboratory comparison of battery impedance analyzers calibration

Year: 2023
Type: article
Abstract: The paper reports a results of series of interlaboratory comparisons of low impedance measurements at frequencies relevant for electrochemical impedance spectroscopy (EIS) of commercial lithium-ion cells. Two comparisons are presented. The first, bilateral comparison has focused on low impedance standards calibration in a full complex plane using digital sampling setups. The second comparison has focused on calibration and use of commercial 4-terminal battery EIS meters. Both comparisons have covered the impedance range from 50μ℧ to 100m℧ across the full complex plane in a frequency range from 0.01Hz up to 5kHz. Finally, the paper summarizes practices identified as critical for achieving measurement compatibility among various labs.
Source: Measurement
Language: English
Open Access status: hybrid
Estimated APC paid: $3,500
OpenAlex ID: W4380201615
FWCI: 0.705
Cites: 10
Cited by: 10