When Impedance Trend Lies, Remote Monitoring Helps for Diagnosis
Year: 2023
Type: article
Study design: Case Report
Abstract: We described the fracture of an ICD lead in a patient at the third device replacement.Through the remote monitoring electric interference was detected on the defibrillator lead despite the pacing and shock impedances during the leadlifetime remained stable and in a normal range. X-ray confirmed the lead failure and patient underwent to a lead replacement.
Source: Annals of Clinical Cardiology
Institution: Ospedale Sandro Pertini
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
OpenAlex ID: W4402115380