Work

Fast Degaussing Procedure for a Magnetically Shielded Room

Year: 2024
Type: article
Abstract: A demagnetization study was conducted on a magnetically shielded room (MSR) at Fraunhofer IPM, designed for applications such as magnetoencephalography (MEG) and material testing. With a composite of two layers of mu-metal and an intermediate aluminum layer, the MSR must provide a residual field under 5 nT for the successful operation of optically pumped magnetometers (OPMs). The degaussing process, employing six individual coils, reached the necessary residual magnetic field within the central 1 m3 volume in under four minutes. Due to the low-frequency shielding factor of 100, the obtained average residual field is shown to be limited by environmental residual field changes after degaussing and not by the degaussing procedure.
Source: Materials
Language: English
Open Access status: gold
Estimated APC paid: $2,955
OpenAlex ID: W4404960106
FWCI: 0.793
Cites: 12
Cited by: 5
Related to: 10