Comparison in humidity (dew-point temperature high range) dew point temperature +30 °C to 95 °C
Year: 2024
Type: article
Abstract: Main text The report summarises the results of the EURAMET.T-K8 key comparison (EURAMET Project No. P 717) in the field of humidity. The comparison covers the dew-point temperature range from +30 °C up to + 95 °C dew point temperature. The report gives details of the scheme, procedure and analysis of the comparison. The overall results display a very good agreement of the 16 participating laboratories. The two dew-point hygrometers used as travelling standard were sufficiently stable. The linkage to the corresponding CCT-K8 comparison will be calculated when the final results for both comparisons are available. Possible procedures are described in the text. In the conclusion, some remarks for improvement for future RMO key comparisons are given. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Source: Metrologia
Authors: Regina Deschermeier, Stephanie Bell, Robert Benyon Puig, Anders Bonde Kentved, Paul Carroll, Vito Fernicola, Jamie García Gallegos, Martti Heinonen, Domen Hudolkin, Rafał Jarosz, Evmorfia Kokkini, Dubhaltach Mac Lochlainn, Helmut Mitter, Seda Oğuz Aytekin, Matthijs Panman, Remo Senn, Danijel Šestan, Denis Smorgon, Mark Stevens, Davor Zvizdić
Institutions: Physikalisch-Technische Bundesanstalt, Instituto Nacional de Técnica Aeroespacial, FORCE Technology (Denmark), National Physical Laboratory, Istituto Nazionale di Ricerca Metrologica, National metrology institute VTT MIKES, VTT Technical Research Centre of Finland, University of Ljubljana, Central Office of Measures, Hellenic Institute of Metrology, National Standards Authority of Ireland, E+E Elektronik, Tübitak National Metrology Institute, VSL National Metrology Institute, Swiss Federal Institute of Metrology, University of Zagreb, Faculty of Mechanical Engineering and Naval Architecture in Zagreb, Nuclear Plasma Laboratories (United States)
Language: English
Subfield: Instrumentation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W4405570652