Work

Comparison of gravimetric standards for hydrogen refuelling stations

Year: 2025
Type: article
Abstract: The EURAMET.M.FF-S15 comparison was organized to validate existing gravimetric standards used for calibrating hydrogen refuelling stations and allow participants to declare CMCs for this new quantity. Coriolis mass flow meters were used as transfer standards. Four laboratories participated between March 2023 and October 2023; CESAME-EXADEBIT (France), METAS (Switzerland), TÜV SÜD NEL (UK) (hereby referred to as NEL) and Justervesenet (Norway). CESAME EXADEBIT is the pilot laboratory. Only two participants could take data with both transfer standards and one with only one transfer standard. Comparison Reference Values were determined for each transfer standard and the degree of equivalence was calculated for each laboratory.
Language: English
Subfield: Instrumentation
Open Access status: gold
Estimated APC paid: $1,910
OpenAlex ID: W4406136970
FWCI: 0.509
Cites: 3
Cited by: 1