Abstract: Intracranial pressure (ICP) monitoring is the mainstay of treatment for patients with brain tumors, hydrocephalus, cerebral edema, and traumatic brain injury (TBI). At present, there are 2 main monitoring methods: non-invasive and invasive. Although invasive monitoring of ICP is the criterion standard, non-invasive approaches based on measurement of optic nerve sheath diameter (ONSD) are becoming increasingly popular because of their simplicity, low cost, and accuracy. Ultrasound (US) measurement of ONSD is especially useful because it allows dynamic evaluation of ICP over the hospital and treatment period in a real-time manner and can be performed bedside for better patient acceptance. This paper reviews the research progress, and highlights the advantages, performance, and results of various ONSD measurement approaches, including US-based ONSD measurement, with regard to its principles, operation method, and clinical applications, as well as ONSD measurement by other techniques such as computed tomography (CT) and magnetic resonance imaging (MRI). New developments, including use of the ONSD/ocular transverse diameter (ETD) index and possible future studies on ONSD-based/assisted therapy and treatment for patients with TBI, stroke, and other conditions that can increase ICP will also be discussed.