Abstract: BACKGROUND: Right ventricular placement of the OmniaSecure defibrillation lead has demonstrated high defibrillation success and low occurrence of lead-related major complications. With the advent of left bundle branch area pacing (LBBAP) as a physiological pacing option, the use of the OmniaSecure lead may be a viable alternative for LBBAP in the implantable cardioverter-defibrillator population. OBJECTIVE: This study aimed to understand the defibrillation efficacy at implant and the safety of the OmniaSecure lead for LBBAP. METHODS: The global Lead Evaluation for Defibrillation and Reliability LBBAP trial enrolled patients indicated for de novo implantation of a primary or secondary prevention implantable cardioverter-defibrillator or cardiac resynchronization therapy with a defibrillator. Per study protocol, LBBAP includes electrocardiographic characterization of left bundle branch pacing and deep septal pacing. The primary efficacy objective was defibrillation efficacy at implant exceeding a prespecified threshold of 88% in ≥160 patients. The primary safety endpoint was to demonstrate the OmniaSecure lead-related major complication rate at 3 months after implant. RESULTS: ). A total of 293 patients were successfully implanted per protocol (95.4%). The primary efficacy objective was met with 100% of patients defibrillated to protocol. The observed freedom from OmniaSecure lead-related major complications at 3 months was 97.9% (95% confidence interval 95.8-100). There were zero study lead fractures, no oversensing adverse events, and stable electrical parameters through 2.6 ± 2.5 months' follow-up. CONCLUSION: These results demonstrate that implantation of OmniaSecure lead when intended for LBBAP results in high defibrillation success at implant, stable short-term electrical parameters, and a low rate of OmniaSecure lead-related major complications at 3 months.
Source: Heart Rhythm
Language: English
Field: Medicine
Open Access status: hybrid
Estimated APC paid: $4,790
OpenAlex ID: W4415321908
FWCI: 5.875
Cites: 29
Cited by: 10
Related to: 10