Navigating the Lead Paradox: Successful Co‐Implantation of Cardiac Contractility Modulation Device and a Micra Leadless Pacemaker
Year: 2026
Type: article
Study design: Case Report
Abstract: BACKGROUND: Wild-type transthyretin cardiac amyloidosis (ATTRwt-CA) can lead to refractory heart failure. A "lead paradox" occurs when patients with a Micra leadless pacemaker require lead-based Cardiac Contractility Modulation (CCM) therapy. CASE SUMMARY: We detail the first co-implantation of CCM and Micra devices in a 78-year-old male with ATTRwt-CA and NYHA III heart failure. A multi-view fluoroscopic technique ensured spatial separation, while specific device programming mitigated electrical crosstalk post-procedure. CONCLUSION: At 1 year, the patient stabilized to NYHA Class II without further hospitalizations. This dual-device strategy is a feasible and safe technical roadmap for this complex clinical problem.
DOI: 10.1111/jce.70244
Institutions: Ospedale San Luigi Gonzaga, University of Turin
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: hybrid
Estimated APC paid: $4,500