Development and experimental validation of a neutron detector system for accelerator-based BNCT irradiation fields
Year: 2026
Type: article
Abstract: Accurate evaluation of neutron leakage dose plays a critical role in radiation protection for accelerator-based boron neutron capture therapy (AB-BNCT). This study reports the experimental validation of a neutron detector system for direct measurement of the neutron ambient dose, H *, in AB-BNCT irradiation fields. The system comprises an S-detector for detecting low-energy neutrons and an F-detector for higher-energy neutrons, with sensitivity extending up to 28 MeV. Detector geometries were optimized using Monte Carlo simulations to reproduce the fluence-to- H * conversion coefficients recommended in the ICRP Publication 116. Prototype detectors were tested in standard neutron fields, including thermal, monoenergetic, and fast neutron sources. The measured detector responses agreed well with calculated values, with deviations ranging from −13% to +4% for response and from −25% to −2% for dose response relative to ICRP values. These results demonstrate the suitability of the developed detector system for evaluating ambient neutron dose in AB-BNCT environments.
Institutions: Kyoto University, Institute for Integrated Radiation and Nuclear Science, Kyoto University, Osaka Medical and Pharmaceutical University, Institute of Laser Engineering, The University of Osaka, Nagoya University, The University of Tokyo, Tohoku University, Research Center for Neutrino Science, Tohoku University, National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology
Language: English
Subfield: Radiation
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $2,210
OpenAlex ID: W7147533332
