Abstract: We present a report on the fabrication and characterization of chromium–nickel atomic force microscopy (AFM) probes with a chromium-coated tip and a bi-metallic cantilever. The introduction of a chromium layer enhances tip hardness and wear resistance, while nickel improves adhesion and reduces brittleness, enabling higher probe usability. The fabrication process was optimized through careful substrate polishing, photoresist selection, and stress control in chromium deposition, which limited the maximum chromium thickness to ∼1 μm. Mechanical testing confirmed that the probes operate reliably in AFM, achieving imaging quality comparable to commercial silicon probes, though resolution was constrained by the relatively large tip radius. Importantly, the design allows for tailoring of stiffness and resonance frequency by adjusting nickel thickness, broadening the applicability of the probes for both contact and intermittent-contact AFM modes. These results establish a proof-of-concept for durable metallic AFM probes engineered for tribological and high-load applications.