Reduction of Klebsiella pneumoniae bacterial load after active disinfection of venous catheter connectors

Year: 2026
Type: article
Abstract: OBJECTIVE: to verify the effectiveness of active disinfection in reducing the bacterial load of Klebsiella pneumoniae inoculated in two types of needleless connectors. METHOD: an in vitro experimental study was carried out with 84 experiments. The disinfection treatments used were sterile gauze with 70% ethanol, non-sterile gauze with 70% ethanol, sterile gauze with Incidin®, and a single-use device with 70% isopropanol. Data were analyzed using the Kruskal-Wallis test and Conover's post-hoc test. RESULTS: all disinfection methods were effective. In valved connectors, the greatest reduction in bacterial load was with 70% ethanol (96.56%), and in extension connectors, the reduction was greater than 99% in all disinfection methods. CONCLUSIONS: the reduction of microbial load in venous access connectors is an important indicator for the prevention of bloodstream infections, in addition to contributing to the promotion of safe practices during nursing care.